of  quantification  of  50,  15  and  50  ng/mL,  respectively.  The  RSDs  of  intra-day  and  inter-day  precision  tests  for  all  three  analytes
          were all less than 10%, with accuracy (relative errors) ranging from -10.00% to 12.96%. The average extraction recovery ranged
          from 92.54% to 100.95%, and the average matrix effect was 89.98%-101.49%. Dilution reliability covered all dilution factors used
          in  the  test  samples,  and  the  absolute  values  of  relative  errors  in  stability  tests  were  all≤16.25%.  The  plasma  concentrations  of
          venetoclax,  busulfan  and  voriconazole  in  enrolled  patients  were  496.20-4  250.45,  233.48-2  002.28  and  475.51-5  710.18  ng/mL,
          respectively. CONCLUSIONS The LC-MS/MS method established in this study is rapid, sensitive and easy to operate, and can be
          used for the therapeutic drug monitoring of venetoclax, busulfan and voriconazole.
          KEYWORDS     venetoclax; busulfan; voriconazole; acute myeloid leukemia; LC-MS/MS; therapeutic drug monitoring


              急性髓系白血病(acute myeloid leukemia,AML)是            1 材料
          最常见的白血病类型之一。据统计,2019 年我国 AML                        1.1 主要仪器
          的发病率约为每10万人口1.24例,占所有急性白血病患                             CalQuant-S 型 LC-MS/MS 系统购自杭州凯莱谱科
          者数的 54% 。目前,异基因造血干细胞移植(alloge‐                      技股份有限公司;EX125DZH型十万分之一天平购自奥
                     [1]
          neic  hematopoietic  stem  cell  transplantation,allo-HSCT)  豪斯仪器(上海)有限公司;5804R型高速冷冻离心机购
                                              [2]
          被视为用于 AML 关键且有效的治疗手段 。白消安在                          自德国 Eppendorf 公司;VORTEX-5 型涡旋混合器购自
          allo-HSCT 预处理中具有关键作用,可有助于清除骨                        海门市其林贝尔仪器制造有限公司。
            [3]
          髓 。然而,该药的血药浓度治疗窗窄且个体差异较                             1.2 主要药品与试剂
                                                                  维奈克拉对照品(批号 271193,纯度 99%)、维奈克
            [4]
          大 ,故临床有必要对接受allo-HSCT预处理的患者体内
                                                              拉 -D8 对 照 品(维 奈 克 拉 的 内 标 ,批 号 288297,纯 度
          白消安的暴露量进行监测,以防止疗效不佳(血药浓度
                                                              95%)、白消安-D8对照品(白消安的内标,批号 263340,
          过低所致)或发生不良反应(剂量过高所致)。另一种常
                                                              纯度 99%)均购自美国 MCE 公司;伏立康唑对照品(批
          用于AML患者的治疗药物是维奈克拉。该药是一种高
                                                              号CHB-F-80,纯度98%)、泊沙康唑对照品(伏立康唑的
          度选择性的 B 细胞淋巴瘤 2 抑制剂 。已有多项研究表
                                        [5]
                                                              内标,批号CHB240124,纯度98%)均购自成都克罗玛生
          明,对于接受 allo-HSCT 治疗的特定人群(如老年、高危
                                                              物技术有限公司;白消安对照品(批号 SB8770,纯度
          或复发/难治性AML患者),在移植前采用维奈克拉联合                          98%)购自北京索莱宝生物技术有限公司。乙腈、乙酸
          白消安+磷酸氟达拉滨预处理方案显示出可接受的安全                            铵、甲酸、二甲基亚砜(DMSO)均为色谱纯,其余试剂为
          性和潜在生存获益         [6―8] 。然而,维奈克拉的体内代谢受               分析纯,水为超纯水。
          多种因素影响,包括饮食差异和与细胞色素 P450 3A4                        1.3 空白血浆
                                                        [9]
         (cytochrome P450 3A4,CYP3A4)抑制剂的联合使用 ;                   空白血浆来自本课题组成员职工体检剩余样本。
          此外,AML患者大多免疫功能低下,发生真菌感染的风                           2 方法与结果
          险很高,特别是侵袭性真菌病,故多接受伏立康唑治                             2.1 色谱与质谱条件
                                                                                      ®
            [10]
          疗 。伏立康唑是一种 CYP3A4 抑制剂,可能导致维奈                            以 Phenomenex  Kinetex   C18 (2.1  mm×50  mm,2.6
          克拉体内暴露量显著增加;此外,维奈克拉药品说明书                            µm)为色谱柱;以 0.1% 甲酸溶液(含 2 mmol/L 乙酸铵)
          指出,当患者联用伏立康唑时,临床可能需要调整维奈                            为流动相A、0.1%甲酸乙腈溶液为流动相B进行梯度洗
          克拉剂量。因此,对患者进行治疗药物监测(therapeutic                     脱(0~0.50 min,15%B;0.50~0.60 min,15%B→28%B;
          drug monitoring,TDM)以精准调整药物剂量十分必要。                  0.60~1.35 min,28%B→58%B;1.35~1.55 min,58%B→
              目前,已有研究者采用液相色谱-串联质谱(liquid                      98%B;1.55~2.59 min,98%B;2.59~2.61 min,98%B→
                                                              15%B;2.61~3.10 min,15%B);流速为0.8 mL/min;柱温
          chromatography-tandem  mass  spectrometry,LC-MS/MS)
                                                              为40 ℃;进样量为5 μL;运行时间为3.10 min。
          法分别检测人血浆中维奈克拉              [11―12] 、白消安 或伏立康
                                                [13]
                                                                  采用电喷雾离子源,以多反应监测(multiple reac‐
            [14]
          唑 的浓度。然而,这些方法存在预处理步骤复杂、灵
                                                              tion monitoring,MRM)方式进行正离子扫描;离子喷射
          敏度低以及不能同时定量3种药物等不足。基于此,本
                                                              电压为 5 500 V;电离温度为 550 ℃;气帘气压力为 25
          研究建立了一种可用于同时测定AML患者血浆中维奈
                                                              psi;雾化气温度为50 ℃;辅助加热气压力为55 psi;碰撞
          克拉、白消安和伏立康唑浓度的 LC-MS/MS 法,并将其                       气压力为 4 psi;用于定量分析的离子对及其余 MRM 参
          用于临床[研究方案经中山大学第七附属医院(深圳)医                           数见表1。
          学 伦 理 委 员 会 批 准 ,批 件 号 为 KY-2025-325-01、KY-         2.2 相关溶液的制备
          2024-412-02],以期为 AML 患者的精准治疗提供 TDM                  2.2.1 对照品溶液和质控溶液
          技术手段,为了解药物相互作用,确保患者安全、合理用                               精密称取维奈克拉对照品5.0 mg,置于5 mL容量瓶
          药奠定基础。                                              中,加 DMSO 溶解并定容,制成质量浓度为 1 mg/mL 的


          · 1324 ·    China Pharmacy  2026 Vol. 37  No. 10                            中国药房  2026年第37卷第10期